Front-Panel Key Reference
Meas Setup
Other
Antenna, Cable, and Other correction factors are generally entered as
NOTE
positive values. This indicates a loss in the external device. User
correction factors are typically entered as negative values which
indicate a gain in the external device.
138
Key Access:
Corrections
Allows you to clear all data from the
Delete Correction
selected amplitude-correction set. The
prompt If you are sure, press key
again to delete will appear on the
display. Pressing
delete the correction set.
Key Access:
Corrections
Allows you to correct for gain or loss other than for
antenna, cable, or user.
Key Access:
Meas Setup
Turns the amplitude correction
Correction On Off
function on or off for the selected set.
The corrections state must be set to On
for the correction to be applied.
Accesses menu keys that allow you to
Edit
create and edit an amplitude-correction
factor set. It puts the analyzer into a
split-screen mode where the correction
data is displayed in a table under the
trace data. Pressing
menu will exit the menu and remove
the table from the screen. New points
will be applied only after the editor is
closed. The Tab keys are very useful for
navigation between rows in the
corrections table.
Key Access:
Corrections
Point
Allows you to create or edit an
amplitude-correction factor data point.
Up to 200 points may be defined for
each set. Enter the point number to be
created or edited by using the numeric
keypad, then press
,
Meas Setup
More,
,
,
Cable
Edit
again will
Delete
,
Meas Setup
More
,
Cable
,
,
More
Corrections
while in this
ESC
,
Meas Setup
More,
,
Other
, or use the
Enter
,
Chapter 2