Agilent Technologies EMC Series User Manual page 136

9 khz- 1.5/3.0/6.7/13.2/26.5 ghz
Hide thumbs Also See for EMC Series:
Table of Contents

Advertisement

Front-Panel Key Reference
Meas Setup
Antenna, Cable, and Other correction factors are generally entered as
NOTE
positive values. This indicates a loss in the external device. User
correction factors are typically entered as negative values which
indicate a gain in the external device.
136
Turns the amplitude correction
Correction On Off
function on or off for the selected set.
The corrections state must be set to On
for the correction to be applied.
Key Access:
Corrections
Accesses menu keys that allow you to
Edit
create and edit an amplitude-correction
factor set. It puts the analyzer into a
split-screen mode where the correction
data is displayed in a table under the
trace data. Pressing
menu will exit the menu and remove
the table from the screen. New points
will be applied only after the editor is
closed. The Tab keys are very useful for
navigation between rows in the
corrections table.
Key Access:
Corrections
Point
Allows you to create or edit an
amplitude-correction factor data point.
Up to 200 points may be defined for
each set. Enter the point number to be
created or edited by using the numeric
keypad, then press
knob, tab, or step keys to move to an
existing point. Press
errors. After selecting a point,
Frequency
Key Access:
Corrections
Frequency
Allows you to enter the frequency value
for an amplitude-correction point.
Enter the frequency value by using the
numeric keypad. Change the frequency
,
,
Meas Setup
More
,
Cable
while in this
ESC
,
Meas Setup
More,
,
Cable
, or use the
Enter
to correct
Bk Sp
becomes active.
,
Meas Setup
More,
,
,
Cable
Edit
Chapter 2