Probes Environmental Requirements; Time And Manpower Requirements - GE Versana Balance Basic Service Manual

Table of Contents

Advertisement

2-2-5 Probes environmental requirements

2-2-5-1 Operation, storage and transport temperatures for probes
CAUTION
Table 2-6:
Temperature
Humidity
Pressure
CAUTION
CAUTION
NOTE:

2-2-6 Time and manpower requirements

Versana Balance – Basic Service Manual
5808768-100 English Rev.9
General Ultrasound system requirements
Probes should be operated, stored, or transported within the
parameters outlined below.
Ensure that the probe face temperature does not exceed the
normal operation temperature range.
Probe Environmental Requirements
Operational
10° - 40° C
50° - 104° F
30- 80%
non-condensing
700 - 1060hPa
Check the room temperature before you use the probe.
Ensure that the probe face temperature does not exceed the
normal operation temperature range.
Refer to
Table 2-1 on page 2-3
settlement time.
Site preparation takes time. Begin site preparation checks as
soon as possible, if possible, six weeks before delivery, to allow
enough time to make any changes.
Storage
-5° - 50° C
23° - 122° F
10 - 90%
non-condensing
700 - 1060hPa
to determine the needed
System view
Transport
-5° - 50° C
23° - 122° F
10 - 90%
non-condensing
700 - 1060hPa
2-11