GE Versana Balance Basic Service Manual page 305

Table of Contents

Advertisement

7-5-5-9 FRU Status
Element
Name
Status
Lower Limit Warnings
Upper Limit Warnings
Lower Limit Errors
Upper Limit Errors
Versana Balance – Basic Service Manual
5808768-100 English Rev.9
FRU Status displays a summary of the voltages for the FRUs
being monitored and the details for each FRU.
To access this page, under System Status, select Voltage and
then select FRU Status.
Figure 7-20. FRU Status
This table shows all the elements available on FRU Status with
descriptions.
Table 7-6:
Name of the FRU.
Status of the voltage for the FRU. Valid values are:
• Normal - Indicates that the fan is within the allowable range.
• Warning - Indicates that the voltage is close to the limit of the allowable range.
• Error - Indicates that the voltage is outside the allowable range.
Number of low limit voltage warnings.
Number of high limit voltage warnings
Number of low limit voltage errors
Number of high limit voltage errors
For more information, see:
7-5-5-8 'Voltage' on page 7-26
7-5-5 'Home' on page 7-16