Electrical Requirements - GE Versana Balance Basic Service Manual

Table of Contents

Advertisement

Site Preparations

2-2-3 Electrical requirements

2-2-3-1 General requirements
NOTE:
NOTE:
2-2-3-2 Electrical requirements for the Ultrasound system
Table 2-3:
Voltage
100-240 VAC
2-2-3-3 Site power outlets
2-6
System view
GE requires a dedicated power and ground for the proper
operation of its Ultrasound equipment. This dedicated power
shall originate at the last distribution panel before the Ultrasound
system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not
shared with any other circuit), and a full size ground wire from
the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not
shared with any other circuit, and a full size ground wire from the
distribution panel to the Ultrasound outlet.
Please note that image artifacts can occur, if at any time within
the facility, the ground from the main facility's incoming power
source to the Ultrasound system is only a conduit.
In the table below, the electrical specifications for the Ultrasound
system includes monitor and on board peripherals.
Electrical Specifications for Versana Balance system
Tolerance
Power Consumption
±10%
Max. 350VA
A dedicated AC power outlet must be within reach of the
Ultrasound system without extension cords. Other outlets
adequate for the external peripherals, medical and test
equipment needed to support this Ultrasound system must also
be present within 1 m (3.2 ft.) of the Ultrasound system.
Electrical installation must meet all current local, state, and
national electrical codes.
Versana Balance – Basic Service Manual
5808768-100 English Rev.9
Frequency
50/ 60HZ