INFINITY CENTRAL STATION & M300 QUICK REFERENCE GUIDE

Transfer of Data

Note: Only trend and demographic data will transfer. Units of measure for
both devices must match. Some devices/software levels do not fully support
electronic data transfer. Only patients in Standby mode are candidates for
transfer. Refer to individual device IFU's.
TRANSFER M300 TO M300
1.
Place the M300 (containing the data to be transferred)
in Standby mode.
2.
Proceed to the parameter box of the admitted M300 you want to
transfer the data to.
3.
Select the Admit/Standby... button on the Bed-View menu bar.
4.
Select the Data Transfer tab.
5.
Select the arrow button next to the Care unit field to select the
desired care unit in which the M300 is located. A list of patients
in the selected care unit appears.
6.
Select the desired patient from the selection list.
7.
Select the Start transfer button. A confirmation pop-up window
appears.
8.
Select Yes in the confirmation popup to start the data transfer.
The message Transfer completed appears after the data have
been transferred.
TRANSFER DRÄGER BEDSIDE MONITOR TO M300
1.
Place the source bedside monitor in Standby mode.
2.
Select admitted M300 parameter box at the ICS to open BedView
page.
3.
Select the Admit/Standby... button on the BedView menu bar.
4.
Select the Data Transfer tab to access the Transfer page.
5.
Select the arrow button next to the Care unit field to select the
desired care unit in which the bedside monitor is located. A list
of patients in the selected care unit appears.
6.
Select the desired patient from the selection list.
7.
Select the Start transfer button. A confirmation pop-up window
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION