Important information about disinfecting
If the product is used by more than one person, using a conventional disinfectant is required.
Only use colourless water-based disinfectants. Observe the instructions for use provided by the manufacturer.

8 Maintenance and repair

8.1 Maintenance

WARNING
Insufficient maintenance
Severe user injuries, damage to the product due to failure to observe maintenance intervals
► Only carry out the maintenance tasks described in this section. All other maintenance and service tasks may
only be carried out by qualified personnel.
► The functionality and operating safety of the product must be verified and a service performed at least
once per year.
► For users with a changing anatomy (for example body dimensions, weight) or users with a changing clinical
picture, have the product inspected, adjusted and serviced at least once every six months.
WARNING
Lack of maintenance
Severe user injuries, damage to the product due to maintenance errors
► Check the adjustment functions for visible signs of damage at least 1 x per month and ensure all screw con­
nections are tight.
WARNING
Improper maintenance
Severe injury of the user due to screw connections coming loose
► Check the tightness of the screw connections at regular intervals.
► Contact the qualified personnel promptly if defects are noted.
The function of the product should be checked before each use.
The product may not be used if defects are noted. This applies in particular in case of instability of the product
or altered driving characteristics as well as problems with the user's seating position or the stability of the seat.
Inform the qualified personnel promptly for the rectification of defects.
This also applies if loose, worn, bent or damaged components, cracks in the frame or broken frame compon­
ents are identified.
Some maintenance tasks can be carried out to a certain extent by the user at home (see Page 39, section
"Maintenance intervals" and see Page 40, section "Maintenance tasks").
In addition, the manufacturer recommends regular maintenance by qualified personnel every 12 months.
Failure to maintain the product can lead to injuries for the user of the product.

8.1.1 Maintenance intervals

The functions described below must be checked by the user or an attendant at the specified intervals:
Component
Main frame
Seat frame
Interface
adapter/seat plate
Angle adjustment
Back support
Dino 3
Task
Check that screw connections are tight
Check that screw connections are tight
Check attachment to the seat frame/back support
Check that removable seating solutions are firmly
locked
Check the swivel axis for ease of operation
Check that screw connections for angle adjust­
ment are tight
Check that the seat tilt locks securely
Check attachment for firm fit
Check that release bolts engage fully
Maintenance and repair
Before each
Weekly
use
X
X
X
X
X
Monthly
X
X
X
X
39