Charging The Prosthesis Battery; Display Of The Current Charge Level; Display Of Battery Charge Level Without Additional Devices - Otto Bock Genium X3 Series Instructions For Use Manual

Bionic prosthetic system
Hide thumbs Also See for Genium X3 Series:
Table of Contents

Advertisement

Available languages

Available languages

2) Connect the round, three-pin plug of the power supply to the receptacle on the battery char­
ger (see fig. 2) so that the plug locks into place.
INFORMATION: Ensure correct polarity (guide lug). Do not use force when connecting
the cable plug to the battery charger.
3) Plug the power supply unit into the outlet (see fig. 3).
→ The green LED on the back of the power supply lights up.
→ The LED ring (status indicator) on the rear of the charger lights up green to indicate the
correct connection to the power supply.
→ If the green LED on the power supply and the LED ring on the battery charger do not light up,
there is an error (see Page 100).

8.2 Charging the prosthesis battery

INFORMATION
When the Protective Cover is installed, the battery charger cable has to point to the upper clos­
ure. A correct knee joint charging process is only ensured with this alignment.

8.3 Display of the current charge level

8.3.1 Display of battery charge level without additional devices

INFORMATION
The charge level cannot be displayed during the charging process, e.g. by turning the prosthes­
is over. The product is in charging mode.
74
1) Connect the inductive charger to the receiver of the charging unit
on the rear of the product. The charger is held in place by a mag­
net.
→ The LED ring on the rear of the charger pulsates purple
(4-second cycle).
→ If the LED ring lights up in a different colour, this indicates an
error (see Page 100).
2) The charging process starts.
→ Once the product battery is fully charged, all LEDs on the side
of the battery charger light up.
3) After the charging process is complete, hold the prosthesis still and
remove the inductive charger from the receiver.
→ A self-test is performed. The joint is operational only after cor­
responding feedback (see Page 104).
1) Turn the prosthesis by 180° (the sole of the foot must face up).
2) Hold still for 2 seconds and wait for beep signals.