Otto Bock Genium X3 Series Instructions For Use Manual page 108

Bionic prosthetic system
Hide thumbs Also See for Genium X3 Series:
Table of Contents

Advertisement

Available languages

Available languages

Interference
immunity tests
Conducted HF inter­
ference according to
IEC 61000-4-6
Radiated HF interfer­
ence according to
IEC 61000-4-3
Note 1: The higher frequency range applies at 80 MHz and 800 MHz.
Note 2: These directives may not be applicable in all cases. The propagation of electromagnetic
factors is influenced by absorption and reflection by buildings, objects and people.
108
IEC 60601 test level
3  V
150  kHz to 80 
eff
MHz
3 V/m 80 MHz to
2.5 GHz
Compliance level
Not applicable
12 V/m 80 MHz to
1 GHz
3 V/m 1 GHz to
2.5 GHz
Electromagnetic
environment direct­
ive
Portable and mobile
wireless equipment
should not be used at
a lesser distance from
the product 'Genium
X3 - Bionic Prosthetic
System 3B5-X3 /
Genium X3 3B5-2'
including the lines
than the recommen­
ded safety distance
calculated using the
equation applicable for
the transmission fre­
quency.
Recommended
safety distance:
d = 0.3 √P for 80 MHz
to 800 MHz
d = 2.3 √P for
800 MHz to 2.5 GHz
where P is the nominal
output of the transmit­
ter in watts (W)
according to the trans­
mitter manufacturer
and d is the recom­
mended safety dis­
tance in metres (m).
The field strength of
stationary radio trans­
mitters at all frequen­
cies according to an
on-site investigation
should be less than
the compliance level.
Interference is pos­
sible in the vicinity of
devices that bear the
following symbol.
a

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

3b5-x33b5-x3-stGenium x3 3b5-2Genium x3 3b5-2-st

Table of Contents