The field strength of stationary transmitters such as base stations of radio telephones and
mobile land radio equipment, amateur radio stations, AM and FM radio and television stations
cannot be precisely determined theoretically in advance. A site survey is recommended to estab­
lish the electromagnetic conditions as a result of stationary HF transmitters. If the field strength
determined at the site of the product 'Genium X3 - Bionic Prosthetic System 3B5-X3 / Genium
X3 3B5-2' exceeds the compliance level specified above, the product 'Genium X3 - Bionic Pros­
thetic System 3B5-X3 / Genium X3 3B5-2' has to be monitored in regards to normal operation at
each place of use. If unusual performance characteristics are noted, taking additional steps may
be necessary, for example changing the orientation or moving the product 'Genium X3 - Bionic
Prosthetic System 3B5-X3 / Genium X3 3B5-2'
Table 6 – Recommended safe distances between portable and mobile HF telecommunica­
tion equipment and the product 'Genium X3 - Bionic Prosthetic System 3B5-X3 / Genium
X3 3B5-2' – for non-life-sustaining devices and systems
Directives and manufacturer's declaration – Recommended safe distances between
portable and mobile HF telecommunication equipment and the product 'Genium X3 -
Bionic Prosthetic System 3B5-X3 / Genium X3 3B5-2'
The product 'Genium X3 - Bionic Prosthetic System 3B5-X3 / Genium X3 3B5-2' is designed for
operation in an electromagnetic environment where HF interference is monitored. The customer
or user of the product 'Genium X3 - Bionic Prosthetic System 3B5-X3 / Genium X3 3B5-2' can
help prevent electromagnetic interference by complying with the minimum distances between
wearable and mobile HF communication equipment (transmitters) and the product 'Genium X3 -
Bionic Prosthetic System 3B5-X3 / Genium X3 3B5-2', as below according to the maximum out­
put of the communication equipment.
Nominal output of
transmitter [W]
0.01
0.1
1
10
100
For transmitters with no maximum nominal output specified in the table above, the recommen­
ded safety distance d in metres (m) can be determined using the equation in the respective
column, where P stands for the maximum nominal output of the transmitter in watts (W) accord­
ing to the transmitter manufacturer's information.
Note 1: The higher frequency range applies at 80 MHz and 800 MHz.
Note 2: These directives may not be applicable in all cases. The propagation of electromagnetic
factors is influenced by absorption and reflection by buildings, objects and people.
Safety distance [m] by transmission frequency
150 kHz to 80 MHz
N/A
N/A
N/A
N/A
N/A
80 MHz to 800 MHz
d = 0.3 √P
0.03
0.09
0.3
0.95
3
800 MHz to 2.5 GHz
d = 2.3 √P
0.23
0.73
2.3
7.3
23
109