Replacing the power supply

Time needed: 5 minutes or less
Tools needed: None
Note: To order parts, contact us.
Warning:
To avoid a shock hazard, do not attempt to remove the cover of the power
supply.
Warning:
Before you replace the power supply, always make sure you have turned off your
instrument and disconnected the power supply from the wall outlet or power strip.
1.—Turn off your instrument.
2.—Disconnect the old power supply.
Note: Turn the locking ring on the power supply connector until you feel it snap into place.
3.—Connect the new power supply.
Simply reverse the order that you disconnected the power cords in the previous step.
4.—Turn on your instrument.
The Power and Scan indicators on the main cover should light and your instrument should
function normally when you turn on the power. If your instrument does not function
normally, turn off the power and check the cable connections between the power supply,
the instrument, and the wall outlet or power strip. If the connectors are seated properly
and your instrument still does not function normally, contact us for assistance.
20
Using Your Nicolet iS10 Spectrometer
Thermo Fisher Scientific