Operational Features - Beckman Coulter Avanti J-26S XP Instructions For Use Manual

High-performance centrifuge
Hide thumbs Also See for Avanti J-26S XP:
Table of Contents

Advertisement

Description
Specifications
Specifications
Temperature
Acceleration
Deceleration
a. During transient conditions, such as acceleration and deceleration, rotor temperature may be outside this range. Refer to
applicable rotor manuals for specific rotor operating range information.
b. To reach temperatures above ambient, the centrifuge is dependent on the frictional heat generated inside the chamber
during operation. At low run speeds or low ambient temperatures, the centrifuge may not be able to achieve some higher
temperatures.

Operational Features

Specifications
Door
Rotor Chamber Diameter
Friction Reduction
System (FRS)
1-8
Description
• Setting range:
— –10 to +40°C (in 1°C increments)
• Accuracy:
— rotor temperature controlled to within 2°C of set temperature (after
a
equilibration)
• Ambient temperature range:
— 16 to 38°C (60 to 100°F)
• Cooling fluid:
— refrigerant 404A (HFC)
maximum or slow
maximum, slow, or off
Description
6.1-cm (2.4-in.) thick plastic resin with steel plate
51.3 cm (20 in.)
190 mm (7.5 in.) Hg
b
PN B10087AB

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents