12.
Connect the inlet fitting on the regulator tee to the output fitting on the Desiccant Tower using the
tubing a nut and a ferrule. Make sure the tube is inserted all the way into the fitting on the
Desiccant Tower before tightening the knurled nut.
13.
If there are no additional instruments to be connected skip to step 17. If there are additional
instruments to be connected, repeat steps 2 - 6 and 14 - 16 for each instrument.
14.
Apply a thin coating of High Vacuum grease to two compression ferrules.
15.
Cut a piece of tubing to connect the regulator tee on the previous instrument to the regulator tee
on the next instrument.
16.
Connect the previous Carrier Gas regulator tee to the next Carrier Gas regulator tee using two nuts
and ferrules.
17.
Using the plug removed in step one, cap the open port on the last regulator tee assembly.

Setting the Gas Supply Pressure

Set the Gas Supply Pressure as follows:
1.
Set the Carrier Gas cylinder or main line supply regulator pressure to 35 psi.
2.
Set the pressure on the Carrier Gas regulator-tee so that the gauge reads 29 psi.
Caution:
The maximum Carrier Gas pressure to the instrument must not exceed 32 psi. Input
pressures greater than 32 psi (2.2 Bar) will damage the system.
Note: When using compressed gas cylinders, the tank must be replaced when the pressure falls below 300 psi (on
the first stage of the regulator).

System Outgassing

MOCON, Inc.
Revision G
Setting Up
2-9