The Leveled Sine Wave Function - Fluke 522A/6 Operator's Manual

Multi-product calibrator
Table of Contents

Advertisement

5522A
Operators Manual

The Leveled Sine Wave Function

The Leveled Sine Wave (LEVSINE) function uses a leveled sine wave, whose amplitude
remains relatively constant over a range of frequencies, to check the oscilloscope's
bandwidth. When you check your oscilloscope, you change the wave's frequency until
the amplitude displayed on the oscilloscope drops 30 %, which is the amplitude that
corresponds to the -3 dB point. Default values are 30 mV p-p, 50 kHz.
To access the LEVSINE menu, press the softkey under MODE until "levsine" appears.
You can press the MODE softkey to cycle through the functions in the order shown, or
you can press P  t o return directly to the SCOPE menu.
Each option in the LEVSINE menu is described below:
10-16
Figure 10-2. Tunnel Diode Pulser Connections
5522A
CALIBRATOR
gjh037.eps
gjh056.eps