How To Set Ac Current Output - Fluke 522A/6 Operator's Manual

Multi-product calibrator
Table of Contents

Advertisement

How to Set AC Current Output

Complete the following procedure to set an ac current output at the AUX or 20A
terminals. If you make an entry error, press G to clear the display, then reenter the
value.
1. Press R to clear any output from the Calibrator,
2. Connect the UUT as described earlier in this chapter under "Connecting the
Calibrator to a UUT."
3. Set the UUT to measure ac current on the desired range.
4. Press the numeric keys and decimal point key to enter the desired current output
(maximum six numeric keys). For example, 123.456.
5. Press a multiplier key, if necessary. For example, press c.
6. Press A.
7. The Control Display now shows the amplitude of your entry. For example,
123.456 mA (below).
8. Press the numeric keys and decimal point key to enter the desired frequency output
(maximum five numeric keys). Press a multiplier key, if necessary. For example,
press the kilo multiplier key K. Then press the H key. For example,
1.1234 kHz (below).
9. Press E. The calibrator clears your entry from the Control Display and copies it
into the Output Display (below is typical).
10. Press O to activate the calibrator output.
Front Panel Operation
How to Set Output
nn079f.eps
nn080f.eps
nn081f.eps
4
4-23