Remote Commands; Commands - Fluke 522A/6 Operator's Manual

Multi-product calibrator
Table of Contents

Advertisement

5522A
Operators Manual
This menu allows editing of the delta () amplitude of both the volts and the current in
the single (sag or swells) mode.
5. To edit the delta () volts part of this waveform, press the V/V blue softkey. At
6. To edit the delta () current part of this waveform, press the I/I blue softkey. At
7. Press the RAMP-UP blue softkey and select 1 second, then
8. Press the WIDTH blue softkey and enter 3, then
Assuming the calibrator is in Operate Mode and a remote *TRG command is received,
the calibrator will delay for 3 seconds and then begin ramping the output down, over a
1.0 second interval to 7.5 Volts and 0.5 Amps. The output will remain at this level for
5 seconds after which it will return immediately to the nominal value, in this case 10 V
and 1 A.

Remote Commands

This section documents the IEEE-488/RS-232 remote commands for the PQ option.
Remote commands duplicate activities that can be initiated from the front panel in local
operation. The following is an alphabetical listing of the PQ option remote commands
with examples. For more information on using remote commands, see Chapter 5,
"Remote Operation".
SEC (or Secondary) refers to the output of the AUX jacks when using dual
outputs from the 5522A. When a dual output is in use, the NORMAL jacks
will be the PRI (or Primary) output and the AUX jacks will be the SEC
(secondary) output. If the AUX jacks are used for single outputs, they are
the PRI output.

Commands

CHIEC <OUTPUT CH.>, <IEC WAVEFROM NUMBER>
(IEEE-488, RS-232, Coupled)
Set composite harmonic wave to a preset IEC wave.
<OUTPUT CH.> is PRI or SEC
<IEC WAVEFROM NUMBER> is 1 or 2
Example: CHIEC PRI,2 ---- sets primary output wave to IEC waveform number 2.
______________________________________________________________________
11-22
the prompt, enter -25 percent, then press
the prompt, enter -50 percent, then press
, and finally
, and finally
.
Note
nn366f.eps
.
.
.

Advertisement

Table of Contents
loading

This manual is also suitable for:

5522a5522a-pq5522a-pq/1g5522a/1ghz5522a-pq/6

Table of Contents