How to Set Resistance Output

Complete the following procedure to set a synthesized resistance output at the Calibrator
front panel NORMAL terminals. If you make an entry error, press G to clear the
display, then reenter the value.
1. Press R to clear any output from the Calibrator.
2. Connect the UUT as described earlier in this chapter under "Connecting the
Calibrator to a UUT."
Since this is a synthesized output, be sure the terminal connections from the
Calibrator to the UUT are LO to LO and HI to HI.
3. Set the UUT to measure resistance on the desired range.
4. Press the numeric keys and decimal point key to enter the desired resistance output
(maximum six numeric keys). For example, 12.3456.
5. Press a multiplier key, if necessary. For example, press K.
6. Press Q.
7. The Control Display now shows the amplitude of your resistance entry. For example,
12.3456 kΩ (below).
8. Press E. The calibrator clears your entry from the Control Display and copies it
into the Output Display (below is typical).
9. Press O to activate the calibrator output.
The softkeys allow selection of three lead-compensation settings and ohms zero.