5522A
Operators Manual

Capacitance

Range
220 to
399.9 pF
0.4 to
1.0999 nF
1.1 to
3.2999 nF
3.3 to
10.9999 nF
11 to
32.9999 nF
33 to
109.999 nF
110 to
329.999 nF
0.33 to
1.09999 μF
1.1 to
3.29999 μF
3.3 to
10.9999 μF
11 to
32.9999 μF
33 to
109.999 μF
110 to
329.999 μF
0.33 mF to
1.09999 mF
1.1 to
3.29999 mF
3.3 to
10.9999 mF
11 to
32.9999 mF
33 to
110 mF
[1]
The output is continuously variable from 220 pF to 110 mF.
[2]
Specifications apply to both dc charge/discharge capacitance meters and ac RCL meters. The maximum allowable peak voltage is
3 V. The maximum allowable peak current is 150 mA, with an rms limitation of 30 mA below 1.1 μF and 100 mA for 1.1 μF and
above.
The maximum lead resistance for no additional error in 2-wire COMP mode is 10 Ω.
[3]
1-16
Absolute Uncertainty,
tcal ±5 °C
[1] [2] [3]
±(% of output + floor)
90 days
1 year
0.38 + 10 pF
0.5 + 10 pF
0.38 + 0.01 nF
0.5 + 0.01 nF
0.38 + 0.01 nF
0.5 + 0.01 nF
0.19 + 0.01 nF
0.25 + 0.01 nF
0.19 + 0.01 nF
0.25 + 0.01 nF
0.19 + 0.01 nF
0.25 + 0.01 nF
0.19 + 0.3 nF
0.25 + 0.03 nF
0.19 + 1 nF
0.25 + 1 nF
0.19 + 3 nF
0.25 + 3 nF
0.19 + 10 nF
0.25 + 10 nF
0.30 + 30 nF
0.40 + 30 nF
0.34 + 100 nF
0.45 + 100 nF
0.34 + 300 nF
0.45 + 300 nF
0.34 + 1 μF
0.45 + 1 μF
0.34 + 3 μF
0.45 + 3 μF
0.34 + 10 μF
0.45 + 10 μF
0.7 + 30 μF
0.75 + 30 μF
1.0 + 100 μF
1.1 + 100 μF
Charge-Discharge Rate
Resolution
Min and Max to
Meet
Specification
0.1 pF
10 Hz to 10 kHz
0.1 pF
10 Hz to 10 kHz
0.1 pF
10 Hz to 3 kHz
0.1 pF
10 Hz to 1 kHz
0.1 pF
10 Hz to 1 kHz
1 pF
10 Hz to 1 kHz
1 pF
10 Hz to 1 kHz
10 pF
10 to 600 Hz
10 pF
10 to 300 Hz
100 pF
10 to 150 Hz
100 pF
10 to 120 Hz
1 nF
10 to 80 Hz
1 nF
0 to 50 Hz
10 nF
0 to 20 Hz
10 nF
0 to 6 Hz
100 nF
0 to 2 Hz
100 nF
0 to 0.6 Hz
10 μF
0 to 0.2 Hz
Allowed Frequency or
Typical Max for
Typical Max for
<0.5 % Error
<1 % Error
20 kHz
40 kHz
30 kHz
50 kHz
30 kHz
50 kHz
20 kHz
25 kHz
8 kHz
10 kHz
4 kHz
6 kHz
2.5 kHz
3.5 kHz
1.5 kHz
2 kHz
800 Hz
1 kHz
450 Hz
650 Hz
250 Hz
350 Hz
150 Hz
200 Hz
80 Hz
120 Hz
45 Hz
65 Hz
30 Hz
40 Hz
15 Hz
20 Hz
7.5 Hz
10 Hz
3 Hz
5 Hz

Advertisement