Fluke 522A/6 Operator's Manual page 255

Multi-product calibrator
Table of Contents

Advertisement

Table 7-11. Verification Tests for Thermocouple Simulation
Output, °C
TC Type
0.00 °C (0.0000 mV)
100.00 °C (1.0000 mV)
-100.00 °C (-1.0000 mV)
10 μV/°C
1000.00 °C (10.0000 mV)
-1000.00 °C (10.0000 mV)
10000.00 °C (100.0000 mV)
-10000.00 °C (-100.0000 mV)
Table 7-12. Verification Tests for Thermocouple Measurement
TC Type
0.00 °C (0.0000 mV)
10000.00 °C (100.0000 mV)
10 μV/°C
-10000.00 °C (-100.0000 mV)
30000.00 °C (300.0000 mV)
-30000.00 °C (-300.0000 mV)
Lower Limit, mV
Input, mV
-0.0030
0.99696
-1.00304
9.99660
-10.0034
99.9930
-100.0070
Lower Limit, °C
-0.30
9999.30
-10000.70
29998.50
-30001.50
Maintenance
Performance Tests
Upper Limit, mV
0.0030
1.00304
-0.99696
10.00340
-9.9966
100.0070
-99.9930
Upper Limit, °C
-0.30
10000.70
-9999.30
30001.50
-29998.50
7
7-17

Advertisement

Table of Contents
loading

This manual is also suitable for:

5522a5522a-pq5522a-pq/1g5522a/1ghz5522a-pq/6

Table of Contents