AC Voltage (Sine Wave) (cont.)
Range
Frequency
10 mV to
10 Hz to 20 Hz
329.999 mV
20 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 5 kHz
5 kHz to 10 kHz
10 kHz to 30 kHz
0.33 V to
10 Hz to 20 Hz
3.29999 V
20 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 5 kHz
5 kHz to 10 kHz
10 kHz to 30 kHz
3.3 v to 5 v
10 Hz to 20 Hz
20 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 5 kHz
5 kHz to 10 kHz
[1]
There are two channels of voltage output. The maximum frequency of the dual output is 30 kHz.
Note
Remote sensing is not provided. Output resistance is <5 mΩ for outputs ≥0.33 V. The AUX output resistance is <1 Ω. The maximum
load capacitance is 500 pF, subject to the maximum burden current limits
Absolute Uncertainty,
tcal ±5 °C
[1]
±(% of output + μV)
90 days
1 year
AUX Output
0.15 + 370
0.2 + 370
0.08 + 370
0.1 + 370
0.08 + 370
0.1 + 370
0.15 + 450
0.2 + 450
0.3 + 450
0.4 + 450
4.0 + 900
5.0 + 900
0.15 + 450
0.2 + 450
0.08 + 450
0.1 + 450
0.07 + 450
0.09 + 450
0.15 + 1400
0.2 + 1400
0.3 + 1400
0.4 + 1400
4.0 + 2800
5.0 + 2800
0.15 + 450
0.2 + 450
0.08 + 450
0.1 + 450
0.07 + 450
0.09 + 450
0.15 + 1400
0.2 + 1400
0.3 + 1400
0.4 + 1400
Introduction and Specifications
Max Distortion and
Max
Resolution
Burden
1 μv
5 mA
10 μv
5 mA
100 μv
5 mA
Detailed Specifications
Noise
10 Hz to 5 MHz
Bandwidth
±(% output + floor)
0.2 + 200 μV
0.06 + 200 μV
0.08 + 200 μV
0.3 + 200 μV
0.6 + 200 μV
1 + 200 μV
0.2 + 200 μV
0.06 + 200 μV
0.08 + 200 μV
0.3 + 200 μV
0.6 + 200 μV
1 + 200 μV
0.2 + 200 μV
0.06 + 200 μV
0.08 + 200 μV
0.3 + +200 μV
0.6 + 200 μV
1
1-13