Pulse Response Calibration With A Tunnel Diode Pulser; The Leveled Sine Wave Function - Fluke 5502A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5502A:
Table of Contents

Advertisement

5502A
Operators Manual

Pulse Response Calibration with a Tunnel Diode Pulser

You can use the Calibrator to drive a tunnel diode pulser. This lets you see pulse edge
rise times as fast as 125 ps.
The Calibrator sources a maximum pulser drive signal of 100 V p-p at 100 kHz. The
recommended (and default) output setting is 80 V p-p at 100 kHz.
Do the subsequent procedure to use a tunnel diode pulser:
1. Connect the Calibrator, tunnel diode pulser, and oscilloscope as shown in Figure 9-2.
2. With the SC600 Option in EDGE mode, push the TDPULSE softkey to "on".
3. Push .
4. Rotate the control on the pulser box to the minimum setting necessary to trigger a

The Leveled Sine Wave Function

The Leveled Sine Wave (Levsine) function uses a leveled sine wave, whose amplitude
remains relatively constant over a range of frequencies, to examine the bandwidth of the
oscilloscope. When you examine your oscilloscope, you change the wave's frequency
until the amplitude shown on the oscilloscope drops 30 %, which is the amplitude that
corresponds to the -3 dB point.
To access the Levsine menu, push the softkey below MODE until "levsine" shows.
9-16
reading.
Figure 9-2. Tunnel Diode Pulser Connections
5502A
CALIBRATOR
gvx037.eps

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents