Calibrator to a UUT" by adapting the voltage and current connections.
3. Set the UUT to measure dc power on the desired range.
4. Press the numeric keys and decimal point key to enter the desired voltage output
(maximum seven numeric keys). For example, 123.4567.
At voltage outputs of 100 volts and above (nominal), you may notice a
slight high-pitched sound. This is normal.
5. Press I to select the polarity of the voltage (default is +).
6. Press a multiplier key, if necessary. For example, press c.
7. Press V.
8. The Control Display now shows the amplitude of your entry. For example,
123.4567 mV (below).
9. Press the numeric keys and decimal point key to enter the desired current output
(maximum six numeric keys). For example, 234.567.
10. Press I to select the polarity of the current (default is +).
11. Press a multiplier key, if necessary. For example, press c.
12. Press A.
13. The Control Display now shows the amplitude of your entries. For example,
123.4567 mV and 234.567 mA (below).
14. Press E. The calibrator clears your entry from the Control Display and copies it
into the Output Display (below is typical).
15. Press O to activate the calibrator output. When changing power output levels, you
must reenter both voltage and current (in either order).
(Enter voltage or current and then a watts entry value using bA. The remaining
volts or current value is calculated and displayed.)
Note
Front Panel Operation
How to Set Output
nn071f.eps
nn082f.eps
nn083f.eps
4
4-25