Fluke 5320A Getting Started Manual

Fluke 5320A Getting Started Manual

Multifunction electrical tester calibrator
Hide thumbs Also See for 5320A:

Advertisement

Quick Links

®
5320A
Multifunction Electrical Tester Calibrator
Getting Started Manual
PN 2634331
September 2006, Rev. 1, 7/07
© 2006, 2007 Fluke Corporation, All rights reserved. Printed in USA.
All product names are trademarks of their respective companies.

Advertisement

Table of Contents
loading

Summary of Contents for Fluke 5320A

  • Page 1 ® 5320A Multifunction Electrical Tester Calibrator Getting Started Manual PN 2634331 September 2006, Rev. 1, 7/07 © 2006, 2007 Fluke Corporation, All rights reserved. Printed in USA. All product names are trademarks of their respective companies.
  • Page 2 Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price.
  • Page 3 Immediately upon arrival, purchaser shall check the packing container against the enclosed packing list and shall, within thirty (30) days of arrival, give Fluke notice of shortages or any nonconformity with the terms of the order. If purchaser fails to five notice, the delivery shall be deemed to conform with the terms of the order.
  • Page 5 OPERATOR SAFETY SUMMARY WARNING HIGH VOLTAGE is used in the operation of this equipment LETHAL VOLTAGE may be present on the terminals, observe all safety precautions! To avoid electrical shock hazard, the operator should not electrically contact the output hi or sense hi binding posts. During operation, lethal voltages of up to 1100 V ac or dc may be present on these terminals.
  • Page 6 Power Source The 5320A is intended to operate from a power source that will not apply more than 264V ac rms between the supply conductors or between either supply conductor and ground. A protective ground connection by way of the grounding conductor in the power cord is essential for safe operation.
  • Page 7 FIRST AID FOR ELECTRIC SHOCK Free the Victim From the Live Conductor Shut off high voltage at once and ground the circuit. If high voltage cannot be turned off quickly, ground the circuit. Get Help! Call loudly for help. Call an emergency number. Request medical assistance. Never Accept Ordinary and General Tests for Death Symptoms of electric shock may include unconsciousness, failure to breathe, absence of pulse, pallor, and stiffness, and well as severe burns.
  • Page 9: Table Of Contents

    Ground Bond Resistance Source ..............6 Line/Loop Impedance Source................ 7 Leakage Current Source ................8 RCD (Residual Current Device)..............8 AC/DC Voltage Calibrator (5320A/VLC only) ..........8 Multimeter ..................... 9 Unpacking and Inspection.................. 11 Service Information ................... 12 Placement and Rack Mounting ................12 Cooling Considerations..................
  • Page 10 5320A Getting Started Manual...
  • Page 11 Title Page Standard Equipment ....................11 Line Power Fuses ....................13 Measurement Input Fuses..................14 Line Power Cord Types Available from Fluke ............14 Front-Panel Features ....................16 Display Features..................... 19 Rear-Panel Features ....................20 Function Default Settings..................23 Accessories......................23...
  • Page 12 5320A Getting Started Manual...
  • Page 13 List of Figures Figure Title Page Line Power Cords Types Available from Fluke............. 14 IEC 61010 Measurement Category (CAT) Levels..........15 Power-up Test Results Screen................22 Meter Mode (Reference State) Display..............22...
  • Page 14 5320A Getting Started Manual...
  • Page 15: Introduction

    Getting Started Introduction The Fluke 5320A and 5320A/VLC are Multifunction Electrical Tester Calibrators (hereafter referred to as the Calibrator), providing full calibration and testing of electrical safety testers. Some examples of these testers are: • Megohm meters • Ground bond testers •...
  • Page 16 1.8 kΩ, which are especially suited for loop impedance, line impedance, and ground bond resistance testing. The Calibrator senses and displays Unit Under Test (UUT) test conditions, types of test current, and current levels up to 25 amps. In the 5320A/VLC, an Active Loop Compensation module will compensate for residual resistance when performing loop and line impedance calibrations.
  • Page 17: Instruction Manuals

    • 5320A Users Manual (CD ROM, PN 2634346) Order additional copies of these manuals using the part numbers provided. For ordering instructions, refer to a Fluke catalog or contact a Fluke sales representative. 5320A Getting Started Manual Use this manual for basic getting started information, contacting Fluke, unpacking, and general specifications.
  • Page 18: General Specifications

    5320A Getting Started Manual General Specifications Warm-Up Time ..........30 minutes Specifications Confidence Level..... 99 % Specifications Interval…………………………...1 year Temperature Performance Operating Temperature ........18 to 28 °C Calibration Temperature (tcal) ......23 °C Temperature Coefficient ......... Temperature coefficient for temperature outside of Tcal ±5 °C between +5 °C to +40 °C is 0.1 x /°C...
  • Page 19: High Resistance Source

    Multifunction Electrical Tester Calibrator Electrical Specifications ⎛ ⎞ R = set resistance between 0.5 Ω to 10 kΩ. Uncertainty ............⎛ ⎞ ⎜ ⎜ ⎟ ⎟ ⎜ ⎟ ⎝ ⎠ ⎝ ⎠ Short Mode Nominal resistance........... <50 mΩ Maximum current..........400 mA ac + dc rms Open Mode Nominal resistance...........
  • Page 20: Ground Bond Resistance Source

    5320A Getting Started Manual Resistance Multiplier Adapter (x1000 multiplier) Resistance range ..........350 MΩ to 10 TΩ Uncertainty and Maximum Ratings Maximum Voltage Uncertainty Range Resolution (ac + dc) Peak (tcal ±5 °C) 10000 V 1.0 % + R 350.0 MΩ to 99.99 GΩ...
  • Page 21: Line/Loop Impedance Source

    Multifunction Electrical Tester Calibrator Electrical Specifications Line/Loop Impedance Source Range ..............25 mΩ to 1.8 kΩ Resolution ............16 discrete values Minimum test voltage/current......10 V/10 mA Uncertainty and Maximum Ratings Absolute Maximum Maximum Nominal Uncertainty of Deviation from Continuous Test Short-term Test Current Resistance...
  • Page 22: Leakage Current Source

    Test uncertainty can be influenced by power line voltage instability Substitute Mode..........0.3 % + 2 μA ac + dc rms Active Mode (5320A/VLC only)....... 0.3 % + 1 μA ac + dc rms RCD (Residual Current Device) Trip Current Range: 0.5 X I and 1 X I Mode ........
  • Page 23: Multimeter

    Multifunction Electrical Tester Calibrator Electrical Specifications Uncertainty……………………………………… 0.02 % Settling Time ............. 300 ms to 3 s, depending on output value AC Voltage Uncertainty and Maximum Burden Current Uncertainty Range Resolution Maximum Burden Current ±(% of Reading + mV) 3 – 29.99 V 0.001 V 0.1 % + 9 500 mA...
  • Page 24 5320A Getting Started Manual AC/DC Current Uncertainty Uncertainty Range Resolution ±(% of Reading + mA) 300 mA 0.1 mA 0.15 % + 0.15 1 mA 0.15 % + 1.5 30 A 10 mA 0.30 % + 15 Phantom Power Range ..............0 to 33 kVA Resolution ............
  • Page 25: Unpacking And Inspection

    If you need to reship the Calibrator, use the original container. If it is not available, you can order a new container from Fluke by identifying the Calibrator’s model and serial number. When you unpack the Calibrator, check for all the standard equipment listed in Table 1.
  • Page 26: Service Information

    Each Calibrator is warranted to the original purchaser for a period of one year beginning on the date received. The warranty is located at the front of this manual. Factory-authorized service and technical advice for the Calibrator is available at Fluke Service Centers. A complete list of service centers is available at www.fluke.com.
  • Page 27: Accessing The Fuses

    To avoid electric shock or fire, do not use makeshift fuses or short-circuit the fuse holder. Table 2. Line Power Fuses Line Voltage Selection Fuse Fluke Part No. 115 Volts T4AH250V (5 x 20 mm) 2743488 230 Volts T2AH250V (5 x 20 mm)
  • Page 28: Connecting To Line Power

    Connect the Calibrator to a properly grounded three-prong outlet. Table 4 lists the line power cord types available from Fluke. Table 4. Line Power Cord Types Available from Fluke Type Voltage Fluke Part No.
  • Page 29: Calibrator Safety Class Rating

    Multifunction Electrical Tester Calibrator Calibrator Safety Class Rating Calibrator Safety Class Rating The Calibrator is rated for 1000 V CAT I and 300 V CAT II measurement categories. WCaution To avoid damage to the Calibrator, do not measure mains in CAT III or CAT IV environments.
  • Page 30: Front-Panel Features

    5320A Getting Started Manual Front-Panel Features Table 5 lists the controls and terminals found on the front panel. Table 5. Front-Panel Features 5320A MULTIFUNCTION ELECTRICAL TESTER CALIBRATOR OUTPUT OUTPUT OPER STBY , HI , mA - SENSE 50V PK 1500V PK...
  • Page 31 A group of interconnected menus is called a menu tree. I Output Function Keys Selects the output function. The output functions are: AC/DC Voltage Calibration (5320A/VLC only) Low Resistance Leakage Current High Resistance J Impedance/RCD Keys Selects the impedance and RCD functions.
  • Page 32 5320A Getting Started Manual Table 5. Front-Panel Features (cont.) Item Description The output adjustment controls. If any of these keys are TPUQ pressed or the knob is rotated, a digit on the Output Display becomes highlighted and the output increments or decrements as the knob is rotated.
  • Page 33: Display Features

    Multifunction Electrical Tester Calibrator Display Features Display Features Table 6 lists the different areas of the display and the information contained in them. Table 6. Display Features ehq003.eps Item Description A Input/Output Displays the selected function and their parameters. B Parameters Displays auxiliary measurements and parameters for the selected function.
  • Page 34: Rear-Panel Features

    5320A Getting Started Manual Rear-Panel Features Table 7 lists the items found on the Calibrator’s rear panel. Table 7. Rear-Panel Features METER LEAKAGE CURRENT T32AH 150V T100mAH 250V T4AH 250V SAFETY WARNING DISCONNECT POWER AND SIGNAL LEADS BEFORE REMOVING COVERS.
  • Page 35: Turning On The Calibrator

    The polarity of the neutral and the line wires must be correct for the 5320A to power on. If the “L-N-PE” test fails during the power-on process, the neutral and line wires should be swapped on the plug end of the 5320A power line cord.
  • Page 36 5320A Getting Started Manual The screen in Figure 3 is displayed when the Calibrator completes these tests. ehq020.bmp Figure 3. Power-up Test Results Screen ehq050.bmp Figure 4. Meter Mode (Reference State) Display If the Calibrator detects that the power line is improperly connected, such as line or hot wire crossed with the neutral wire, it will display “Fail.”...
  • Page 37: Warm Up

    Table 9 lists the available accessories for the Calibrator. Table 9. Accessories Model Description 5320CASE Transit case for 5320A Calibrator Y5320 Rack mount kit for mounting a 5320A in a standard 19 inch rack. 5320A-LOAD Load resistor adapter for hipot leakage current calibration...
  • Page 38 5320A Getting Started Manual...