Operators Manual
3. Press the E key to enter the offset and then P.

Editing and Error Output Settings

All Calibrator outputs can be edited using the front panel Edit Field knob and associated
L, W, and e keys. In addition, multiply X and divide D keys edit the
output by decades. The difference between the original output (reference) and edited
output is displayed as an "error" between the two settings. This allows you to edit a value
to achieve a correct reading at the UUT and thereby calculate an error in ±% or ppm
(parts per million) if it is less than ±1000 ppm. Table 4-2 lists the actions that cause the
calibrator to exit the error mode and return to the original reference output, or to output a
new reference, as selected.
E
I or E
A new keypad entry + E
N
X
D
R

How to Edit the Output Setting

When you initially source an output from the Calibrator, you enter a specific value. For
example, 10.00000 V dc. To edit the output value to suit your application, turn the front
panel Edit Field knob clockwise to increase the value or counter-clockwise to decrease
the value. (The Edit Field controls will not operate if you are in any setup function. Press
the P key one or more times to exit a setup function.)
To select a higher order digit, use an Edit Field cursor key L or W. The output digit
in edit is always underlined (see below).
The momentary display of the letter u in the Output Display when editing during OPR
(Operate) indicates "unsettled," that is, the Calibrator output is settling with a new value.
4-48
Table 4-2. Keys That Exit Error Mode
Keys
Returns to the previous reference value.
Establishes a new reference.
Establishes a new reference.
Establishes the present output as a new reference.
Sets the calibrator to ten times the reference value and establishes a new
reference.
Sets the calibrator to one-tenth the reference value and establishes a new
reference.
Returns to the power-up state.
Action
nn114f.eps