Fluke 522A/6 Operator's Manual page 249

Multi-product calibrator
Table of Contents

Advertisement

Table 7-7. Verification Tests for AC Voltage (Normal) (cont.)
Range
Output
32.999 mV
30.000 mV
32.999 mV
30.000 mV
32.999 mV
30.000 mV
329.999 mV
33.000 mV
329.999 mV
33.000 mV
329.999 mV
300.000 mV
329.999 mV
300.000 mV
329.999 mV
300.000 mV
329.999 mV
300.000 mV
329.999 mV
300.000 mV
329.999 mV
300.000 mV
329.999 mV
300.000 mV
329.999 mV
300.000 mV
329.999 mV
300.000 mV
3.29999 V
0.33000 V
3.29999 V
0.33000 V
3.29999 V
3.00000 V
3.29999 V
3.00000 V
3.29999 V
3.00000 V
3.29999 V
3.00000 V
3.29999 V
3.00000 V
3.29999 V
3.00000 V
3.29999 V
3.00000 V
3.29999 V
3.00000 V
3.29999 V
3.00000 V
3.29999 V
3.29000 V
32.9999 V
3.3000 V
32.9999 V
3.3000 V
32.9999 V
30.0000 V
32.9999 V
30.0000 V
32.9999 V
30.0000 V
32.9999 V
30.0000 V
32.9999 V
30.0000 V
Frequency
Lower Limit
50 kHz
29.970 mV
100 kHz
29.898 mV
450 kHz
29.770 mV
45 Hz
32.987 mV
10 kHz
32.987 mV
9.5 Hz
283.350 mV
10 Hz
299.917 mV
45 Hz
299.950 mV
1 kHz
299.950 mV
10 kHz
299.950 mV
20 kHz
299.947 mV
50 kHz
299.902 mV
100 kHz
299.788 mV
500 kHz
299.450 mV
45 Hz
0.32989 V
10 kHz
0.32989 V
9.5 Hz
2.83350 V
10 Hz
2.99920 V
45 Hz
2.99952 V
1 kHz
2.99952 V
10 kHz
2.99952 V
20 kHz
2.99946 V
50 kHz
2.99920 V
100 kHz
2.99823 V
450 kHz
2.99340 V
2 MHz
45 Hz
3.2990 V
10 kHz
3.2990 V
9.5 Hz
28.3350 V
10 Hz
29.9919 V
45 Hz
29.9957 V
1 kHz
29.9957 V
10 kHz
29.9957 V
Maintenance
Performance Tests
Upper Limit
30.030 mV
30.102 mV
30.230 mV
33.013 mV
33.013 mV
316.650 mV
300.083 mV
300.050 mV
300.050 mV
300.050 mV
300.053 mV
300.098 mV
300.212 mV
300.550 mV
0.33011 V
0.33011 V
3.16650 V
3.00080 V
3.00048 V
3.00048 V
3.00048 V
3.00054 V
3.00080 V
3.00178 V
3.00660 V
[1]
0.07500 V
3.3010 V
3.3010 V
31.6650 V
30.0082 V
30.0044 V
30.0044 V
30.0044 V
7