Press Y on the calibrator and switch the DMM function switch to w.
i.
Set the calibrator output to 350 μA at 0 Hz. and press O. Verify the error is
j.
within specification.
k. Repeat the previous step using 3500 μA at 0 Hz.
Press Y on the calibrator and press the blue key on the DMM to switch to ac
l.
measurements.
m. Set the calibrator output to 350.0 μA at 60 Hz and press O. Verify the error is
within specification.
n. Repeat the previous step with the following calibrator settings:
11. Test the High current function.
a. Press R on the calibrator.
b. Verify that the calibrator is in standby and connect the DMM as shown in
Figure 4-20.
87
TRUE RMS MULTIMETER
MIN MAX
RANGE
HOLD
H
REL
Hz
PEAK MIN MAX
mV
mA
V
A
m
A
V
OFF
mA mA
A
COM
V
Figure 4-20. Cable Connections for Testing an 80 Series High Amps Function
c. Set the calibrator output to 3.5 A at 0 Hz and press O. Verify the error is
within specification.
d. Repeat the previous step using 10.0 A at 0 Hz. Verify the error is within
specification.
e. Press Y on the calibrator and press the blue key on the DMM to switch to ac
measurements.
AC Current
Frequency
35.0 mA
350.0 mA
350.0 mA
AC Current
Frequency
350.0 μA
3500.0 μA
3500.0 μA
1.0 kHz
60 Hz
1.0 kHz
1.0 kHz
60 Hz
1.0 kHz
Front Panel Operation
Sample Applications
5522A
CALIBRATOR
gjh027.eps
4
4-59