Time Marker Specifications; Wave Generator Specifications - Fluke 522A/6 Operator's Manual

Multi-product calibrator
Table of Contents

Advertisement

5522A
Operators Manual

Time Marker Specifications

Time Marker into 50 Ω
1-Year Absolute Uncertainty at
Cardinal Points, tcal ± 5 °C
Wave Shape
Typical Output Level
Typical Jitter (rms)
Sequence
Adjustment Range
Amplitude Resolution
[1]
t is the time in seconds.
[2]
Typical rise time of square wave and 20 %-pulse (20 % duty cycle pulse) is <1.5 ns.
Time marker uncertainty is ±50 ppm away from the cardinal points.
[3]

Wave Generator Specifications

Wave Generator Characteristics
Amplitude
Range
1-Year Absolute Uncertainty, tcal
±5 °C, 10 Hz to 10 kHz
Sequence
Typical DC Offset Range
Frequency
Range
Resolution
1-Year Absolute Uncertainty, tcal ± 5 °C
[1]
The dc offset plus the wave signal must not exceed 30 V rms.
10-6
5s to 50 ms
±(25 + t x
[1]
1000 ppm
spike or
square
[2]
>1 V p-p
<10 ppm
5-2-1 from 5 s to 1 ns (e.g., 500 ms, 200 ms, 100 ms )
[3]
At least ±10 % around each sequence value indicated above.
4 digits
20 ms to 100 ns
50 to 20 ns
±2.5 ppm
±2.5 ppm
spike, square, or
spike or square
20 %-pulse
[2]
>1 V p-p
>1 V p-p
<1 ppm
<1 ppm
Square Wave, Sine Wave, and Triangle Wave into 50 Ω or 1 MΩ
into 1 MΩ: 1.8 mV to 55 V p-p
into 50 Ω: 1.8 mV to 2.5 V p-p
±(3 % of p-p output + 100 µV)
1-2-5 (e.g., 10 mV, 20 mV, 50 mV)
0 to ± (≥40 % of p-p amplitude)
10 Hz to 100 kHz
4 or 5 digits depending upon frequency
±(25 ppm + 15 mHz)
10 ns
5 to 1 ns
±2.5 ppm
±2.5 ppm
square or
sine
sine
[2]
[2]
>1 V p-p
>1 V p-p
<1 ppm
<1 ppm
[1]

Advertisement

Table of Contents
loading

This manual is also suitable for:

5522a5522a-pq5522a-pq/1g5522a/1ghz5522a-pq/6

Table of Contents