2. 4. Mains Connection
➀ Power Supply Switch; ➁ Mains Connection
Figure 3: Mains Connection: Fisherbrand accuSpin Micro 17 (left); Fisherbrand accuSpin Micro 17R
(right)
1.
Turn off the power supply switch.
2.
Make sure that the power supply cable specification agrees with the safety standards
of your country.
3.
Make sure that the voltage and frequency are the same as the figures on the nameplate
4.
Make sure that the power supply cable is plugged in properly.
NOT ICE
NOT ICE
Transport and Set Up
Connect the centrifuge into grounded electrical sockets only.
Rated current is doubled during acceleration. Mind this load on the
power supply line.