2. 3. Transporting
Before transporting a centrifuge make sure that
the power supply cable is unplugged and removed from the centrifuge.
ƒ
the rotor is removed.
ƒ
CAUTION Damage to centrifuge or drive shaft due to movement of an installed rotor. Always
remove the rotor before you transport the centrifuge.
the centrifuge door is closed.
ƒ
CAUTION Hands can be squeezed by an open centrifuge door. Always close the centrifuge
door before you transport the centrifuge.
Before transporting a rotor make sure that
all components, like adapters and buckets are removed to prevent dropping
ƒ
damage.
the centrifuge is lifted at both sides and not at the front or back.
ƒ
WARNING Always lift the centrifuge on both sides. Never lift the centrifuge by its front or
back.
Figure 2: Lifting the Centrifuge at Both Sides
Transport and Set Up
35 / 89