4. 6. Autoclaving
Always disassemble all parts before autoclaving, e.g. lids need to be removed before
autoclaving a bucket or rotor.
If not stated otherwise on the parts themselves, all parts can be autoclaved at 121 °C for
20 min. Refer to
"Rotor Specifications" on page 20
is the hematocrit rotor at 134 °C for 20 min. Refer to
for details on rotors.
Make sure that the necessary sterility is achieved according to your requirements.
After autoclaving, treat the entire surface of aluminum parts including the cavities with
corrosion protection oil (70009824).
CAU T IO N
NOT ICE
4. 7. Service
Fisher Scientific recommends having the centrifuge and accessories serviced once a year
by an authorized service technician. The service technician checks the following:
electrical equipment and connections
ƒ
suitability of set-up site
ƒ
centrifuge lid lock and safety system
ƒ
rotor
ƒ
fixation of rotor and drive shaft of the centrifuge
ƒ
protective casing
ƒ
Before service, centrifuge and rotors should be thoroughly cleaned and decontaminated to
ensure full and safe inspection can be completed.
Fisher Scientific offers inspection and service contracts for this work. Any necessary repairs
are performed for free during the warranty period and afterwards for a charge. That is only
valid if the centrifuge has only been maintained by an authorized Fisher Scientific service
technician.
A validation of the centrifuge is recommended and can be ordered from customer service.
Maintenance and Care
Never exceed the permitted temperature and duration when autoclaving.
No chemical additives are permitted in the steam.
for details on rotors. Only exception
"Rotor Specifications" on page 20