Caution
Be sure that the power supply is the
same voltage that is specified on the
refrigerator's data plate.
Warning
For personal safety this unit must be
properly grounded. DO NOT under
any circumstance cut or remove the
ground prong from the power cord.
DO NOT use a plug adapter that
eliminates the ground prong.
Warning
Catalog numbers ending in HK are
international 3-Wire systems and
should not be confused with 230V
domestic units which are 4-Wire sys-
tems. Consult a qualified electrician
before powering this unit.
Caution
Do not use an extension cord. Use of
an ungrounded cord or an overloaded
circuit VOIDS the compressor warran-
ty.

Electrical Connection

Determine the total amount of current presently being
used by other apparatus connected to the circuit that will
be used by this refrigerator. It is critical that this added
current demand and other equipment on this circuit not
exceed the rating of the fuse or circuit breaker in use.
The frequency and nominal voltage requirements for the
unit are specified on the data plate, which is located on
the interior upper left side. Plug the unit only into a power
source that meets these requirements. Low line voltage is
often the cause of service complaints. With the unit run-
ning, check that the line voltage is within ±10% of that
specified on the data plate.
The customer should have the wall receptacle and circuit
checked by a qualified electrician to verify the receptacle
is properly grounded and is connected to the proper
power supply. The following list gives the NEMA
plug/receptacle configuration for the corresponding refrig-
erators.
Laboratory Refrigerators
Catalog No.'s
13-986-120\122
13-986-124\126\128
13-986-128S\130\132
13-986-121\123
13-986-125\127\129
13-986-129S
13-986-131\133
Chromatography Refrigerators
Catalog No.'s
13-986-134
13-986-136\136S\138\140 5-20P
13-986-135\137\137S
13-986-139\141
UNPACKING AND INSTALLATION
NEMA Plug
NEMA Receptacle
5-15P
5-15R
5-15P
5-15R
5-15P
5-15R
L14-20P
L14-20R
L14-20P
L14-20R
6-15P
6-15R
L14-20P
L14-20R
NEMA Plug
NEMA Receptacle
5-20P
5-20R
5-20R
L14-20P
L14-20R
L14-20P
L14-20R