1� 6� 6� Pcr 4 X 8 Rotor - Fisher Scientific Fisherbrand accuSpin Micro Series Instruction Manual

Table of Contents

Advertisement

1. 6. 6. PCR 4 x 8 Rotor
Items Supplied
Item
PCR 4 x 8 Rotor
Table 30: Items Supplied PCR 4 x 8 Rotor
General Technical Data
Maximum Permissible Load
Maximum Number of Cycles
Radius (max. / min.)
Angle
Aerosol-tight
Max. Autoclaving Temperature
Table 31: General Technical Data PCR 4 x 8 Rotor
Rotor Performance Data
17 Series Centrifuges – PCR 4 x 8 Rotor
Voltage
Maximum Speed
Maximum RCF-Value
Acceleration / Braking Time
Sample Heating at max speed, 23 °C
ambient temperature, 60 min running time
21 Series Centrifuges – PCR 4 x 8 Rotor
Voltage
Maximum Speed
Maximum RCF-Value
Acceleration / Braking Time
Sample Heating at max speed, 23 °C
ambient temperature, 60 min running time
Technical Specifications
Article No�
75003440
4 x 4 g (32 x 0.2 g)
50 000
6.6 cm / 4.7 cm
45°
Yes
121 °C
230 V
13 300
13 100
10 s / 11 s
31 °C
230 V
14 800
16 200
12 s / 13 s
33 °C
30 / 89
Quantity
1
120 V
13 300
13 100
10 s / 11 s
31 °C
120 V
14 800
16 200
12 s / 13 s
33 °C