Install The Divider/Oxygen Sensor Assembly - Cardinal Health AirLife nCPAP Hardware Service Manual

Table of Contents

Advertisement

13.5

Install the Divider/Oxygen Sensor Assembly

This assembly contains electronic components that are ESD sensitive.
ESD control and handling measures should be followed while performing
1.
Attach the silicone tubing to the fitting on the Bottom Case.
2.
Connect the Oxygen Sensor Cable to Remote Sensor Board (PCA-02) at CN14
3.
Position the Divider/Oxygen Sensor Assembly into its mounting slots.
4.
Gently insert the Remote Board Bulkhead Cable connector through the
corresponding slot in the Divider/Oxygen Sensor Assembly.
Attach silicone tube (1)
Connect Oxygen Sensor
Cable (2)
© © © © 2008, Cardinal Health
AirLife
CAUTION
this process.
Page 54 of 81
®
nCPAP Driver Service Manual