Cardinal Health AirLife nCPAP Hardware Service Manual page 46

Table of Contents

Advertisement

9.
10. Disconnect the silicone tubing on the Divider/Oxygen Sensor Assembly from the
11. Remove the Divider/Oxygen Sensor Assembly from the Bottom Case.
Disconnect
connector, pinch
and push
through Divider
(1)(2)
Disconnect
luer lock (7)
Disconnect
Oxygen Sensor
wire to PCA-02, (9)
Disconnect silicone
tube from case
(10)
© © © © 2008, Cardinal Health
Disconnect Oxygen Sensor Cable from CN14 on Remote Sensor Board (PCA-02).
barbed fitting on the Bottom Case.
This assembly contains electronic components that are ESD sensitive.
ESD control and handling measures should be followed while performing
CAUTION
this process.
Disconnect Remote
Board connector (5)
Page 46 of 81
®
AirLife
nCPAP Driver Service Manual
LCD Backlight
Power Supply
Remove manifold
mounting screws
(8)
Pinch and push
connector
through Divider
(6)

Advertisement