Install the Top Case

1.
When replacing the Top Case, verify that all assemblies inside the driver are secure.
Check that all connectors are fully latched and all luer fitting connections are snug.
2.
Gently place the Top Case on top of the Bottom Case, ensuring that all cables and
hoses are not pinched or kinked.
3.
Looking from the front of the driver, ensure all metal panels slide into their respective
slots in the Top Case. Lower the Top Case until it's fully seated on the Bottom Case.
4.
Turn the unit upside down.
5.
Insert the four (4) #8 screws into the mounting holes and secure using a 9/64" Allen
wrench. Torque to 5 in-lbs.
Reassembled nCPAP driver with
Top Case removed.
© © © © 2008, Cardinal Health
AirLife
Align metal panels with slots in
Invert unit, install mounting screws