REASSEMBLING THE NCPAP DRIVER

This section provides detailed procedures on replacing components and reassembling the
nCPAP driver. These reassembly procedures are provided in order, starting with an empty
Bottom Case.
13.1

Assemble the Control Knob/Selector Switch

1. Insert the Selector Switch through the keyway in the Bottom Case. Be sure to align
the notch on the switch with the key in the case.
2. Insert the lock washer and the nut. Tighten to 12 in-lbs. using a 7/16" wrench.
3. Position the Control Knob on the post, taking care to align the flats. Press on.
13.2

Install the Oxygen Bleed Fitting

1. Thread Oxygen Bleed Fitting into threaded insert in back left corner of Bottom Case.
2. Torque to 5 in-lb using a 5/16" nut driver.
13.3

Install Battery Assembly

Insert the Battery Assembly into the Bottom Case so that the terminals are closest to the
interior of the case.
Properly installed Battery Assembly.