12.4
Removing the Battery Assembly
The power switch does not disconnect the driver from the main power
supply. This can only be accomplished by disconnecting the AC Adapter
In order to remove the Battery Assembly, the Power Cable Assembly must first be removed.
1.
The Power Cable Assembly is mounted at the rear of the Bottom Case, behind the
Battery Assembly. It is held in place by friction from the foam pads on the Battery
Assembly.
2.
Disconnect negative terminal of Power Cable Assembly from Battery Assembly.
3.
Gently pull the Power Inlet Panel from behind the Battery Assembly.
4.
Disconnect positive terminal of Power Cable Assembly from Battery Assembly.
5.
Disconnect Valve Drive Cable by squeezing the Bulkhead Connector's tabs and
gently pushing it through the slot in the Divider/Oxygen Sensor Assembly.
6.
Gently lift the Battery Assembly from its mount.
© © © © 2008, Cardinal Health
WARNING
from either the driver or the AC outlet.