12.7

Remove the Outlet Fitting Assembly

The power switch does not disconnect the driver from the main power
supply. This can only be accomplished by disconnecting the AC Adapter
1.
Remove Front Panel Assembly, Manifold Assembly, and Divider/Oxygen Sensor
Assembly per instructions in sections 12.3, 12.5, and 12.6.
2.
Use a T-10 Torx driver to loosen and remove the mounting screw that secures the
Outlet Fitting Assembly to the Bottom Case.
This assembly contains electronic components that are ESD sensitive.
ESD control and handling measures should be followed while performing
3.
Slide the Outlet Fitting Assembly out at an upward angle.
4.
Use new cable assemblies if they are provided with the replacement Outlet Fitting
Assembly, otherwise utilize the two (2) Inlet Pressure Board Cables (red/white/black
cables) and the Remote Board to Bulkhead Cable from the replaced assembly.
© © © © 2008, Cardinal Health
WARNING
from either the driver or the AC outlet.
CAUTION
this process.
Page 47 of 81
®
AirLife
nCPAP Driver Service Manual
Remove T-10 Torx screw