Mounting Bracket And Clamp Assembly - Cardinal Health AirLife nCPAP Hardware Service Manual

Table of Contents

Advertisement

11.1

Mounting Bracket and Clamp Assembly

The Mounting Bracket and Clamp Assembly can be installed in two (2) different orientations.
Both are depicted below. Removing the Mounting Bracket and Clamp Assembly is not
necessary to service the driver - instructions are included here only if you wish to change the
orientation.
1. Place nCPAP driver upside down on a clean, dry surface.
2. Use a 9/64" Allen wrench to remove the (3) #8 mounting screws and lock washers.
3. Reorient the bracket to one of the two positions shown below and reattach using the
(3) #8 mounting screws and lock washers.
4. Inspect the rubber feet. Replace worn feet if necessary.
Mounting Bracket and Clamp Assembly,
Rear Installation-Standard Orientation
© © © © 2008, Cardinal Health
AirLife
Mounting Bracket and Clamp Assembly,
Page 32 of 81
®
nCPAP Driver Service Manual
Optional Side Orientation

Advertisement

Table of Contents
loading

Table of Contents