DRIVER MAINTENANCE SCHEDULE

Note: In-depth Preventative Maintenance is to be performed by a competent, fully qualified
service technician or engineer. Preventative Maintenance must be performed at a minimum
frequency per the table below using only approved spare parts.
Required Inspection and Replacement Schedules
Component
Air filter media
(external)
Air and O2 Inlet
Screens
Power Cord/ AC
Adaptor
Oxygen Sensor
Filters (Internal)
Battery Assembly
Manifold Assembly
Divider/Oxygen
Sensor Assembly
System
Oxygen vigorously accelerates combustion. To avoid an explosion hazard,
do not use any instrument or piece of equipment that may have been
© © © © 2008, Cardinal Health
Inspection Schedule
Visual bowl inspection at each start-up
No routine inspection
Visual inspection prior to each start-up;
Electrical inspection annually
No routine inspection
Check annually per section 14
Check bleed flow annually per section 14
Check FiO
Accuracy annually per section 14
2
Perform Comprehensive Performance Test
annually or after any service or repair
exposed to oil or grease contamination.
Page 29 of 81
®
AirLife
WARNING
nCPAP Driver Service Manual
Replacement Schedule
Frequency
Section
12 months
12 months
60 months
12 months
24 months
60 months
60 months
n/a
11.2
11.3
n/a
12.2
12.4
12.5
12.6
n/a

Advertisement