Ethernet Interface Features And Specifications, Socket Interface Specifications - YOKOGAWA WT5000 User Manual

Precision power analyzer communication interface
Hide thumbs Also See for WT5000:
Table of Contents

Advertisement

1.2
Ethernet Interface Features and Specifications,
Socket Interface Specifications
Ethernet Interface Features
Reception Feature
• You can use the reception feature to specify the same settings that you specify by using the touch
panel or the front panel keys.
• Receive output requests for measured and computed data, panel setup parameters, and error
codes.
Transmission Feature
The instrument can (1) transmit measured and computed data, (2) transmit panel setup parameters
and the status byte, and (3) error codes when errors occur.
Ethernet Interface Specifications
Electrical and mechanical specifications
Simultaneous connections
Communication protocol
Connector
Switching between Remote and Local Modes
Switching from Local to Remote Mode
The instrument switches to remote mode when it is in local mode and it receives a :COMMunicate:
REMote ON command from the PC.
Remote Mode
• The REMOTE LED turns on.
• Touch panel operations and front panel key operations except the UTILITY key are disabled.
• The local mode settings are retained even when this instrument switches to remote mode.
Switching from Remote to Local Mode
When the instrument is in remote mode and you press UTILITY, the instrument switches to local mode.
However, this does not work if the instrument has received a :COMMunicate:LOCKout ON command
from the PC. The instrument switches to local mode when it receives a :COMMunicate:REMote OFF
command from the PC, regardless of the local lockout state.
Local Mode
• The REMOTE LED turns off.
• Touch panel operations and front panel key operations are enabled.
• Settings entered in remote mode are retained even when this instrument switches to local mode.
Note
You cannot use the Ethernet interface simultaneously with other interfaces (GP-IB and USB interfaces).
Setting the Timeout Value
If the instrument is not accessed within a given period of time (specified by the timeout value), it will
disconnect from the network. The timeout value can be set to Infinite or in the range of 1 s to 3600 s.
The default value is Infinite.
For instructions on how to set the timeout value, see section 1.4, "Configuring the instrument Ethernet
Settings."
1-2
Complies with IEEE802.3, Auto-MDIX
1
TCP/IP (VXI-11)
RJ-45
IM WT5000-17EN