5.8

Restore factory setting

5.9

Viewing the device status

V3
Use this menu item to restore the factory settings in the device.
With All default the control parameters, internal Pt1000, and other
parameters are reset to factory settings.
With Control parameter only the control parameters are reset to the
factory setting.
With internal Pt1000 only the calibration of the internal temperature
probe is reset to the factory calibration.
With Miscellaneous all other parameters are reset to the factory set-
tings. This includes:
The range of the temperature limit values is reset to 81 °C and 3 °C.
The baud rate is reset to 9600 bauds .
Operating button locking is deactivated.
The temperature control unit is reset to °C .
The sound volume of the signal tone is reset to Loud .
The display brightness is reset to Level 5 .
1.
Select the Menu  Setup  Factory setting menu item.
2.
Select the menu item for the parameter that you want to reset.
3.
Press [OK] to confirm.
4.
Select the [Yes] menu item.
5.
Press [OK] to confirm.
The factory setting stored in the device is restored.
In the "Device status" menu item, you can have the different current tech-
nical data of your device displayed. It isn't possible to set or change values.
You can have the following data displayed:
Error memory
The product line
The device type
The software version
The serial number
The device data with
Temperatures in the device
Pump data
Actuating signal for the heating and cooling
The operating hours.
1.
Select the Menu  Setup  Device status menu item.
2.
Select the menu item for the parameter that you want to view.
3.
Press [OK] to confirm.