2022-01
THE LIQUID LEVEL SENSOR IS UNABLE TO DETECT FOAM;
THEREFORE, IT IS IMPORTANT THAT THE WASTE BOTTLE IS
EMPTIED AS SOON AS THE FOAM LEVEL HAS REACHED THE
MAXIMUM FILLING LEVEL INDICATED ON THE WASTE BOTTLE.
WASTE BOTTLE – FOAMING
WHEN USING WASH BUFFERS THAT SHOW A STRONG TENDENCY
OF FOAMING, EMPTY THE WASTE BOTTLE AS SOON AS THE
FOAM-LEVEL HAS REACHED THE MAXIMUM FILLING LEVEL
INDICATED ON THE WASTE BOTTLE. ADDITIONALLY, ADD A
COMMERCIALLY AVAILABLE ANTI-FOAMING AGENT (SUCH AS
SILICONE OIL) TO THE EMPTY WASTE BOTTLE TO REDUCE
IF FOAMING CONTINUES TO BE A PROBLEM, WE RECOMMEND
SWITCHING TO A LARGER WASTE BOTTLE (NOT PROVIDED WITH
THE HYDROFLEX PLUS) AND ADDITIONALLY INCREASE THE
CONCENTRATION OF ANTI-FOAMING AGENT IN THE WASTE
BOTTLE. TO HELP FACILITATE BREAKDOWN OF THE FOAM IN THE
WASTE BOTTLE, CAREFULLY SWIRL THE WASTE BOTTLE FROM
TIME-TO-TIME TO IMPROVE MIXING BETWEEN FOAM LAYER AND
REFILL ANTI-FOAMING AGENT AFTER EMPTYING WASTE BOTTLE.
FOR EXAMPLE, WHEN USING THE WACKER ANTI-FOAM EMULSION
SE47 (WACKER ARTICLE CODE 21640582), THE RECOMMENDED
CONCENTRATION IS 1ML OF ANTIFOAMING AGENT FOR 1 LITER
FOR ANTIFOAMING AGENTS FROM OTHER MANUFACTURERS, USE
CONCENTRATIONS AS RECOMMENDED BY CORRESPONDING
Liquid Bottle Empty
If the liquid bottle is empty before a program is started, the following message is
displayed:
LLD ERROR
OK
Fill the liquid bottle and press OK to remove the message and start the program
again.
HYDROFLEX PLUS No. 30199906 Rev. No. 1.1
9. Trouble Shooting and Error Messages
WARNING
WASTE BOTTLE – FOAMING
WARNING
FOAMING.
ANTI-FOAMING AGENT.
OF WASTE SOLUTION.
MANUFACTURERS.