PARAMETERS
Dispensing uniformity
Residual volume
Carry over
between wells
Power
Supply
Consumption
Physical
Outside dimensions
Weight
Environmental
Ambient temperature
Operation
Storage
Relative Humidity
Operation
Storage
Others
Overvoltage category
Pollution degree
Method of disposal
2022-01
CHARACTERISTICS
<= 4 % CV
Measured under following conditions:
8-way manifold, 300 µl, dispensing rate 3 (250 µl/s/well), wash buffer, Greiner
96 well flat bottom plate
Measured under following conditions:
<= 2 µl / well (Greiner 96-well flat bottom plate)
-
Crosswise aspiration, 8-way manifold, aspiration rate: 3, aspiration time: 4 s,
head speed: 10 mm/s, wash buffer.
-
Residual volume of <= 2 µl / well cannot be guaranteed if a 16-way manifold is
used and if not, all wells are filled with liquid.
<= 5 µl / well (Greiner 96-well round bottom and V-shaped bottom plates)
-
Single aspiration, 8-way manifold, aspiration rate: 3, aspiration time: 4 s, head
speed: 10 mm/s, wash buffer.
-
Residual volume of <= 5 µl / well cannot be guaranteed if a 16-way manifold is
used and if not, all wells are filled with liquid.
<= 1 ppm
The Carry over between wells <= 1 ppm cannot be guaranteed if not all wells
are filled with liquid.
Auto-sensing
100 - 120 V or 220 - 240 V, 50/60 Hz
< 65 VA
Width: 275 mm, Depth: 366 mm, Height: 180.5 mm
(Width: 10.8 inches, Depth: 14.4 inches, Height: 7.1 inches)
6.6 kg
15°C to 35°C (59°F to 95°F)
-20°C to 60°C (-4°F to 140°F)
20 % to 80 %
non-condensing
10 % to 80 %
non-condensing
II
2
Electronic waste (infectious waste)
HYDROFLEX PLUS No. 30199906 Rev. No. 1.1
2. General
19
Need help?
Do you have a question about the HYDROFLEX PLUS and is the answer not in the manual?