3.

Installation

3.1

Installation Requirements

3.1.1

Required Working Area

3.1.2

Power Requirements

3.2

Unpacking and Inspection

2022-01
When installing, moving, or connecting the instrument, follow the instructions in
this publication. Tecan does not accept the responsibility for injury suffered by
anyone attempting these operations without following the instructions in this
publication, nor for damage incurred to the instrument.
Make sure the laboratory meets all the requirements and conditions described in
this chapter.
Select a location to place the instrument that is flat, level, vibration free, away
from direct sunlight, and free from dust, solvents and acid vapors.
Allow at least 10 cm (4 inches) between the instrument and the wall or any other
equipment. Do not place any items close to the instrument that could obstruct
airflow.
The liquid bottles should be positioned at the same height as the instrument. If
the liquid bottles are placed above or below the instrument (maximum height
difference of 1 meter is allowed), the dispensing pump must be recalibrated by a
service technician.
For information regarding outer dimensions and weight of the instrument, see
chapter 2.3 Instrument Specifications.
The instrument is designed to operate at either 100 - 120 V or 220 - 240 V.
No voltage setting is required as the instrument automatically senses the supplied
voltage. See 2.3 Instrument Specifications for further information.
Connect the instrument only to an electricity supply system with protective earth.
DO NOT REPLACE DETACHABLE MAIN POWER SUPPLY CORDS
WITH INADEQUATELY RATED CORDS.
1. Visually inspect the container(s) for damage before opening.
Report any damage immediately.
2. Place the carton in an upright position and open it.
3. Lift the instrument out of the carton and place it in the selected location.
4. Remove the transport protection from the instrument and manifold.
5. Visually inspect the instrument for loose, bent or broken parts.
Report any damage immediately.
6. Compare the serial number on the rear panel of the instrument with the serial
number on the packing slip.
Report any discrepancy immediately.
7. Check the instrument accessories against the packing list.
HYDROFLEX PLUS No. 30199906 Rev. No. 1.1
CAUTION
3. Installation
25