2022-01
Priming is performed using the following procedure:
PRIME
PRIME
<
<
>
>
OK
OK
PRIME SOLUTION OK?
<
>
YES
WASTE BOTTLE OK?
<
>
YES
VOLUME
<
>
OK
TIME
<
>
OK
CHANNEL 1
-
+
EXIT OK
PROGRAM
<
>
OK
HYDROFLEX PLUS No. 30199906 Rev. No. 1.1
8. Maintenance and Cleaning
Select and confirm PRIME procedure
Confirm rinsing solution and tube in bottle
Confirm waste bottle is empty
Select the duration TIME (5 – 99 s) or the
VOLUME (15 – 820 ml)
Select required channel
(1 - 4, depending on instrument configuration)
Select OK and the prime procedure starts
Instrument is ready to perform wash procedures