16.

Fault finding

 See Section 2.
The OMD-2008 Series will indicate several malfunctions in the status line of the display.
Pressing the "OK" button will display additional information.
Status
Reading
System Alarm LED
Alarm Circuit 1,2
Reason
Recommendations
Status
Reading
System Alarm LED
Alarm Circuit 1,2
Reason
Recommendations
Important Notes
0..15
No Alarm
Low oil content.
-Continue operation.
-Apply standard
maintenance
schedule.
FW!
0..49/ EE
Green/ Blinking
- Freshwater is
enabled.
- Remote fresh
water input is linked
19 -> 20).
- Switch back to
sample stream.
Page: 60 of 102
OK
15..49
Green/ Blinking
Alarm
Normal Operation.
Resume normal
maintenance
schedule.
Flow!
Alarm
-Terminals 15&16
open.
- A magnet is
missing.
- Flow sensor
unsatisfied.
- Magnet sensor
defective.
Check sample flow
rate.
EE
Sample reading is
out of range:
-oil content too
high.
-dirty Sample
Glass Tube.
-Air bubbles in
sample stream.
-Wait until oil
content is within
the range.
-Clean Sample
Glass Tube.
E-VALVE?
Any
Red
EV FW Valve
position not
detected.
(For EV Units only)
Check connectors
and wiring.