13.0 FAULT FINDING

See Section 2 for important notes.
The OMD-2005 will indicate several malfunctions in the status line of the display.
Pressing the "OK" button will lead into an information window, similar to the items
listed in the table below.
Status
Reading
OK
0..49
OK
EE
FW !
0..49 / EE Green /
Sample?
EE
Com?
EE
Datalog?
0..49 / EE Red /
Int.Err
Issue: 27.05.08
DECKMA HAMBURG GmbH
System-Alarm-
Alarm-
circuit
circuit 1,2
LED
Alarm
Green /
No
Normal
Blinking
operation
Green /
No
Alarm
Blinking
No
Alarm
Blinking
Red /
Yes
Alarm
Steady
Red /
Yes
Alarm
Steady
Yes
Alarm
Steady
Red /
Yes
Alarm
Steady
Instruction Manual OMD-2005
Reason
Normal operation
-
Sample reading is out
Wait until oil content is
of range:
within the range,
clean sample tube
Oil content too high,
dirty sample tube
Freshwater is enabled
-
Meter is not able to
Check sample, clean
measure the sample:
sample tube
no water in, oil content
according Page 21
much too high, no light
transmission possible
No communication
Check connection
between computer unit
between computer
and measuring cell
unit and measuring
cell
Datalogging is not
Insert the active
possible:
memory card
no DECKMA card in
Datalogging is not
Insert the active
possible:
memory card
a read only card is in
Datalogging is not
Activate card or insert
possible:
the active memory
card
a new DECKMA card is
in
Internal error
Restart the system
Servicing
Page 20 of 25