4.4
OMD-2008 Piping
Connect the OMD-2008 Monitor to the sample point of the oily-water separator outlet
and to a source of oil free water employing 6 mm OD copper or stainless-steel pipe. The
sampling point should be located on a vertical section of the separator outflow piping to
minimize the effects of any entrained air. The sampling point should be at a level above
the outlet of the monitor to ensure a proper sample flow rate.
Figure 4. Piping OMD-2008
Page: 20 of 102