Deckma Hamburg OMD-2008 Series Instruction Manual page 62

15ppm bilge alarm
Hide thumbs Also See for OMD-2008 Series:
Table of Contents

Advertisement

Status
Reading
System-Alarm-LED
Alarm Circuit 1,2
Reason
Recommendations
Status
Reading
System-Alarm-LED
Alarm Circuit 1,2
Reason
Recommendations
Com?
EE
No communication
between Computer
Unit and Measuring
Cell.
Check
connection
between Computer
Unit and Measuring
Cell.
Int.Err
Red/Steady
Alarm
Internal error.
Restart the system.
Page: 61 of 102
Sample?
Red/Steady
Alarm
Meter is not able to
measure the
sample:
no water in, oil
content much too
high no light
transmission
possible.
-Check sample,
check sample flow
rate, clean Sample
Glass Tube
according to section
12.
OMD-2008
Series Operating
instructions
Desicc.
0..49/EE
Green/ Blinking
Normal Operation
Measuring
Cell
Humidity
critically
high (>40%rH)
Check/
Replace
desiccator.
Data log?
0..49/EE
Data logging is not
possible:
No DECKMA
memory card
inserted.
A read only card has
been inserted.
-A new DECKMA
memory card has
been inserted but
has not been
activated.
Insert the active
memory card.
New memory card.
Activate the
memory card.
Humid
Sample
temperature below
dew point.
Instantaneous
condensation
possible.
Check/
Replace
desiccator.